12. Power on the chassis.

Related Information

Installing the Software

 on page 4

Direct Connections to the NI 5698

The NI 5698 is a precision RF instrument that is sensitive to ESD and transients. Ensure you
take the following precautions when making direct connections to the NI 5698 to avoid
damaging the device.

Caution

 

Apply external signals only while the NI 5698 is powered on. Applying

external signals while the device is powered off may cause damage.

Ensure you are properly grounded when manipulating cables or antennas connected to the
NI 5698 RF IN connector.

If you are using noninsulated devices, such as a noninsulated RF antenna, ensure the
devices are maintained in a static-free environment.

If you are using an active device, such as a preamplifier or switch routed to the NI 5698
RF IN connector, ensure that the device cannot generate signal transients greater than the
RF and DC specifications of the NI 5698 RF IN connector.

Connecting to the NI 5668R

Figure 3.

 NI 5668R with NI 5698 Front Panel Interconnection

0 VDC
+ 20 dBm MAX > 10 MHz
+10 dBM MAX < 10 MHz

NOISE

SOURCE

0/28 VDC

ACCESS

ACTIVE

NI PXIe-5698

RF Preamplifier

10 MHz – 26.5 GHz

RF IN

+17 dBm MAX

±10 VDC

RF OUT

+25 dBm MAX

0 VDC

ACCESS

ACTIVE

ESD

SENSITIVE

ACCESS ACTIVE

LO3 

800 MHz 

LO2 

4 GHz

LO1 

3.2 GHz - 8.3 GHz

OUT

ALL PORTS

50 

Ω

 

NI PXIe-5653

Synthesizer

REF IN

10 MHz 

5 V p-p MAX

REF OUT

10 MHz

1.5 V p-p MAX

REF OUT

 

100 MHz

1.5 V p-p MAX

1

1. Semi-rigid 2.92 mm-to-2.92 mm Coaxial Cable

6

  | 

ni.com

  | 

NI PXIe-5698 Getting Started Guide

Содержание PXIe-5698

Страница 1: ...t may be impaired if it is used in a manner not described in this document Hot Surface If the NI 5698 has been in use it may exceed safe handling temperatures and cause burns Allow the NI 5698 to cool...

Страница 2: ...ry rules Caution To ensure the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance the length of all I O cables mus...

Страница 3: ...river Software DVD 6 NI PXIe 5698 Getting Started Guide This Document 7 Read Me First Safety and Electromagnetic Compatibility 8 Maintain Forced Air Cooling Note to Users Preparing the Environment Ens...

Страница 4: ...ion Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut down or restart later Installing the NI 5698 Cautio...

Страница 5: ...p and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 2 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Downward Unlatched...

Страница 6: ...devices are maintained in a static free environment If you are using an active device such as a preamplifier or switch routed to the NI 5698 RF IN connector ensure that the device cannot generate sign...

Страница 7: ...LL performance and may cause insertion loss Use an appropriate torque wrench or torque screwdriver and 2 92 mm driver bit to ensure all 2 92 mm connections are properly torqued 2 92 mm connectors for...

Страница 8: ...evice Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance The signal pins of this product...

Страница 9: ...written to in order to control the device GREEN The module is ready to be programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for...

Страница 10: ...one device a Select the system device with the appropriate downconverter in the configuration tree b In the Associated Devices section use the RF Conditioning drop down listbox to select the NI 5698 c...

Страница 11: ...In the MAX configuration tree expand Devices and Interfaces 2 Expand the Chassis tree to see the list of installed devices and press F5 to refresh the list 3 If the module is still not listed power o...

Страница 12: ...ignal Analyzers Help NI PXIe 5698 Specifications NI RFSA Soft Front Panel NI RFSA Instrument Driver NI RF Vector Signal Analyzers Help NI RFSA Examples Located using the NI Example Finder Located at n...

Страница 13: ...the DoC for your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration NI corporate headquart...

Страница 14: ...party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain rel...

Отзывы: