Ensure you are properly grounded when manipulating cables or antennas connected to the
NI 5696.
•
If you are using noninsulated devices, such as a noninsulated RF antenna, ensure the
devices are maintained in a static-free environment.
•
If you are using an active device, such as a preamplifier or switch routed to the NI 5696,
ensure no signal transients are sourced to the NI 5696.
Connecting to the NI 5654
Figure 4.
NI 5654 with NI 5696 Front Panel Interconnection
250 kHz–20 GHz Signal Generator
250 kHz–20 GHz Amplitude Extender
ACCESS
ACTIVE
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
RF AMP
OUT
ATTN IN
+30 dBm MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
ALC OUT
±15 VDC MAX
RF OUT
Reverse Power
±30 dBm MAX
RF IN
±15 dBm
MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
ESD
SENSITIVE
ALL PORTS
50
Ω
1
2
1. Semi-Rigid SMB-to-SMB Cable
2. Semi-Rigid SMA-to-SMA Cable, Labeled T
1.
Use the SMB-to-SMB cable to connect the ALC OUT connector on the NI 5696 front
panel to the ALC IN connector on the NI 5654 front panel.
2.
Use Cable T to connect the RF OUT connector on the NI 5654 front panel to the RF IN
connector on the NI 5696 front panel.
NI PXIe-5696 Getting Started Guide
|
© National Instruments
|
7