GETTING STARTED GUIDE

NI PXIe-5694

IF Conditioning Module

Note

 

Before you begin, install and configure your chassis and controller.

This document explains how to install, configure, and test the NI PXIe-5694 (NI 5694). The
NI 5694 is an IF conditioning module. The NI 5694 ships with the NI-RFSA instrument
driver, which you use to program the device.

To access NI 5694 documentation, navigate to 

Start

»

All Programs

»

National Instruments

»

NI-RFSA

»

Documentation

.

Caution

 

The protection provided by this product may be impaired if it is used in a

manner not described in this document.

Hot Surface

 

If the NI 5694 has been in use, it may exceed safe handling

temperatures and cause burns. Allow the NI 5694 to cool before removing it from
the chassis.

Contents

Electromagnetic Compatibility Guidelines...............................................................................2
Verifying the System Requirements......................................................................................... 2
Unpacking the Kit.....................................................................................................................2

Verifying the Kit Contents................................................................................................ 3

Preparing the Environment....................................................................................................... 3
Installing the Software.............................................................................................................. 4
Installing the NI 5694............................................................................................................... 4

Direct Connections to the NI 5694................................................................................... 5
NI 5694 Front Panel..........................................................................................................6

Configuring the NI 5694 in MAX............................................................................................ 8
Troubleshooting........................................................................................................................ 9

Why Is the ACCESS LED Off When the Chassis is On?.................................................9
What Should I Do if the NI 5694 Doesn't Appear in MAX?..........................................10
What Should I Do if the Module Fails the Self-Test?.....................................................10

Where to Go Next................................................................................................................... 11
Worldwide Support and Services............................................................................................ 11

Содержание PXIe-5694

Страница 1: ...r not described in this document Hot Surface If the NI 5694 has been in use it may exceed safe handling temperatures and cause burns Allow the NI 5694 to cool before removing it from the chassis Conte...

Страница 2: ...the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance the length of all I O cables must be no longer than 3 m 10...

Страница 3: ...he environment you are using the NI 5694 in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56 10 to 90...

Страница 4: ...edges or the metal bracket 1 Ensure the AC power source is connected to the chassis before installing the modules The AC power cord grounds the chassis and protects it from electrical damage while yo...

Страница 5: ...is using the front panel mounting screws Note Tightening the top and bottom mounting screws increases mechanical stability and also electrically connects the front panel to the chassis which can impro...

Страница 6: ...N connector ensure that there are no signal transients greater than the RF and DC specifications for the device that are being generated and sourced to the NI 5694 IF IN connector NI 5694 Front Panel...

Страница 7: ...te frequency IF signal to the input connector on a digitizer for example the IF IN connector on the NI PXIe 5622 IF IN Receives a signal from the IF OUT connector of an RF downconverter for example th...

Страница 8: ...may indicate the module has exceeded approved operating temperature and thermal shutdown has occurred or that the module has detected a power supply failure If the power supply fails contact NI techn...

Страница 9: ...le associations may be lost when you move the modules to different chassis slots 6 Self test the device by selecting the device in the configuration tree and clicking Self Test in the MAX toolbar The...

Страница 10: ...start the system 4 Navigate to the Device Manager Operating System Description Windows 8 Right click the Start screen and select All apps Control Panel Hardware and Sound Device Manager Windows 7 Sele...

Страница 11: ...ware EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Example...

Страница 12: ...formation support phone numbers email addresses and current events Refer to the NI Trademarks and Logo Guidelines at ni com trademarks for information on National Instruments trademarks Other product...

Отзывы: