Configuring the PXIe-5665 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which NI hardware products are in the system and how they are
configured. MAX is automatically installed with NI-RFSA.
1. Launch MAX.
2. In the configuration tree, expand
Devices and Interfaces
to see the list of installed NI
hardware.
Note
If you are using the PXIe-5665 with the LabVIEW Real-Time Module,
expand
Remote Systems
. Find your target IP address or name, expand it, and
then expand
Devices and Interfaces
.
Installed modules appear under the name of their associated chassis.
3. Expand your
Chassis
tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note
If you do not see your module listed, press <F5> to refresh the list of
installed modules. If the module is still not listed, power off the system, ensure
the module is correctly installed, and restart.
4. Record the identifier MAX assigns to the hardware. Use this identifier when
programming the PXIe-5665.
5. Associate the hardware modules that comprise your device.
a) Select the PXIe-5603 or PXIe-5605 that is identified as not configured in the
configuration tree.
b) In the Associated Devices section, select the appropriate module from each system
component drop-down listbox.
For the PXIe-5665, you must associate the PXIe-5622 IF digitizer module and the
PXIe-5653 LO source module with the PXIe-5603 or PXIe-5605 RF downconverter.
c) Click
Save
in the MAX toolbar.
Note
Module associations may be lost when you move the modules to
different chassis slots.
6. Self-test the device modules by selecting the modules in the configuration tree, and
clicking
Self-Test
in the MAX toolbar. Repeat this step for all modules in your
PXIe-5665 system.