34
|
ni.com
|
PXIe-5665 Calibration Procedure
24. Close the PXIe-5665 session.
If the frequency response and absolute amplitude accuracy verification procedures determines
that the PXIe-5665 is outside its limits, refer to
Worldwide Support and Services
for information
about support resources or service requests.
Table 14.
Absolute Amplitude Accuracy Verification Test Limits (Preamplifier Enabled)
Frequency
Device
As-Found Limit
(dB)
As-Left Limit
(dB)
*
612.5 MHz
PXIe-5665
3.6 GHz VSA
±0.35
±0.30
PXIe-5665
14 GHz VSA
±0.70
±0.60
10 MHz to
100 MHz
PXIe-5665
3.6 GHz VSA
±1.10
±0.80
PXIe-5665
14 GHz VSA
±1.45
>100 MHz to
2.8 GHz
PXIe-5665
3.6 GHz VSA
±0.80
±0.60
PXIe-5665
14 GHz VSA
±1.15
>2.8 GHz to
3.6 GHz
PXIe-5665
3.6 GHz VSA
±0.80
±0.60
PXIe-5665
14 GHz VSA
±1.20
±0.75
*
section of this document for more information
about as-left limits.
Table 15.
PXIe-5665 14 GHz VSA Absolute Amplitude Accuracy Verification Test Limits
(Preamplifier Disabled, PXIe-5665 14 GHz VSA Preselector Enabled)
Frequency
As-Found Limit (dB)
As-Left Limit (dB)
*
>3.6 GHz to 7.5 GHz
±4.0
±1.5
>7.5 GHz to 8.5 GHz
±4.0
±1.5
>8.5 GHz to 14 GHz
±4.0
±1.5
*
section of this document for more information
about as-left limits.