20
|
ni.com
|
PXIe-5665 Calibration Procedure
3.
Set the RF source 1 and RF source 2 frequency and power values according to the first row
in Table 7.
Table 7.
PXIe-5665 RF Source Power (Combined) Test Settings (Frequencies
>700 MHz)
Start
Frequency
Stop
Frequency
Step Size
(MHz)
RF Source 1
RF Source 2
Frequency
Offset
Power
(dBm)
Frequency
Offset
Power
(dBm)
800 MHz
3.500 GHz
100
+350 kHz
-10
-350 kHz
-10
3598.95 MHz
—
—
+350 kHz
-10
-350 kHz
-10
3.6 GHz
13.8 GHz
200
+350 kHz
-10
-350 kHz
-10
13.99895 GHz
—
—
+350 kHz
-10
-350 kHz
-10
800 MHz
3.500 GHz
100
+350 kHz
-30
-350 kHz
-30
3598.95 MHz
—
—
+350 kHz
-30
-350 kHz
-30
3.6 GHz
—
—
+350 kHz
-30
-350 kHz
-30
800 MHz
1.700 GHz
100
+1 MHz
+9.1
0 Hz
-24
800 MHz
1.700 GHz
100
+1 MHz
+8
0 Hz
-24
800 MHz
1.700 GHz
100
+1 MHz
-13
0 Hz
-54
800 MHz
1.700 GHz
100
+1 MHz
-15
0 Hz
-54
1.800 GHz
14.000 GHz
100
+1 MHz
+7.1
0 Hz
-24
1.800 GHz
14.000 GHz
100
+1 MHz
+6
0 Hz
-24
1.800 GHz
3.600 GHz
100
+1 MHz
-16
0 Hz
-54
1.800 GHz
3.600 GHz
100
+1 MHz
-18
0 Hz
-54