PXIe-5665 Calibration Procedure
|
© National Instruments
|
75
5.
Disable the RF source 1 and RF source 2 outputs.
Note
When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6.
Disable the PXIe-5665 preamplifier.
7.
(PXIe-5665 14 GHz VSA)
Disable the PXIe-5665 preselector.
8.
Set the RF source 2 frequency according to the first row in Table 47 or Table 48 as
appropriate.
9.
Set the RF source 1 frequency 1 MHz above the RF source 2 frequency.
10. Set the PXIe-5665 center frequency according to the values listed in Table 47 or Table 48.
11. Set the RF source 2 power to the
RF Source 2 Programmed Power
from Table 49 or
Table 50 as appropriate.
12. Set the RF source 1 power to the
RF Source 1 Programmed Power
from Table 49 or
Table 50 as appropriate.
Table 47.
PXIe-5665 3.6 GHz Gain Compression Test Frequencies
(Frequencies >700 MHz)
Start Frequency
Stop Frequency
Step Size
800 MHz
3.3 GHz
500 MHz
3.6 GHz
—
—
Table 48.
PXIe-5665 14 GHz Gain Compression Test Frequencies
(Frequencies >700 MHz)
Start Frequency
Stop Frequency
Step Size
800 MHz
3.3 GHz
500 MHz
3.6 GHz
—
—
3.8 GHz
13.8 GHz
500 MHz
14 GHz
—
—