PXIe-5665 Calibration Procedure
|
© National Instruments
|
73
Frequencies >700 MHz
1.
Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the
anti-distortion test fixture, as shown in Figure 27 and Figure 28.
Figure 27.
PXIe-5665 3.6 GHz VSA Gain Compression Verification
(Frequencies >700 MHz) Equipment Setup
1
RF Source 1
2
RF Source 2
3
SMA (m)-to-SMA (m) Cable
4
SMA (m)-to-SMA (m) Cable
5
Anti-Distortion Test Fixture
6
SMA (m)-to-SMA (m) Cable