72
|
ni.com
|
PXIe-5665 Calibration Procedure
22. Compare the
Gain Compression
to the verification test limits in Table 45 or Table 46 as
appropriate.
23. Disable the output of RF source 1 and RF source 2.
Note
When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
24. Close the PXIe-5665 session.
If the gain compression for frequencies
≤
700 MHz
verification procedure determines that the
PXIe-5665 is outside its limits, refer to
Worldwide Support and Services
for information about
support resources or service requests.
Table 45.
Gain Compression Verification Test Limits
(Frequencies
≤
700 MHz, Preamplifier Disabled)
Frequency
Limit
Compression
(dB)
Applied Power
(dBm)
10 MHz to
700 MHz
As-Found
1
+8
As-Left
*
1
+9.1
*
section of this document for more information
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum
applied power. The applied power includes the measurement uncertainty to guarantee the
required minimum applied power.
Table 46.
Gain Compression Verification Test Limits
(Frequencies
≤
700 MHz, Preamplifier Enabled)
Frequency
Limit
Compression
(dB)
Applied Power
(dBm)
10 MHz to
100 MHz
As-Found
1
-18
As-Left
*
1
-16
>100 MHz to
700 MHz
As-Found
1
-15
As-Left
*
1
-13
*
section of this document for more information
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum
applied power. The applied power includes the measurement uncertainty to guarantee the
required minimum applied power.