66
|
ni.com
|
PXIe-5665 Calibration Procedure
23.
(PXIe-5665 14 GHz VSA)
Repeat steps 6 to 21 for test frequencies between 3.8 GHz and
7 GHz with the PXIe-5665 preselector enabled.
24. Repeat steps 6 to 22 for the frequencies in Table 38 with the PXIe-5665 preamplifier
enabled.
25. Compare the
SHI
to the verification test limits in Table 39, Table 40, or Table 41 as
appropriate.
Table 38.
SHI Verification Test Frequencies
Start Frequency (MHz)
Stop Frequency
Step Size
300.0333125
1.3000333125 GHz
200 MHz
1,800
—
—
Table 39.
SHI Verification Test Limits (Preamplifier Disabled,
PXIe-5665 14 GHz VSA Preselector Enabled)
Source
Frequency
Device
As-Found Limit
(dBm)
As-Left Limit
(dBm)
*
300 MHz to
<700 MHz
PXIe-5665
+42
+43
700 MHz to
1.8 GHz
PXIe-5665
3.6 GHz VSA
+50
+51
PXIe-5665
14 GHz VSA
+44
+45
>1.8 GHz to
3.75 GHz
†
PXIe-5665
14 GHz VSA
+54
+56
>3.75 GHz to
4.25 GHz
†
+54
+56
>4.25 GHz to
7 GHz
†
+54
+56
*
section of this document for more information
about as-left limits.
†
Source frequencies >1.8 GHz apply only to the PXIe-5665 14 GHz VSA.