PXIe-5665 Calibration Procedure
|
© National Instruments
|
61
22. Compare the
TOI Minimum
to the verification test limits in Table 34 or Table 35 as
appropriate.
23. Disable the output of the RF source 1 and RF source 2.
Note
When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
24. Close the PXIe-5665 session.
If the third-order intermodulation distortion for frequencies >700 MHz verification procedure
determines that the PXIe-5665 is outside its limits, refer to
Worldwide Support and Services
for
information about support resources or service requests.
Table 34.
TOI (Frequencies >700 MHz) Verification Test Limits (Preamplifier Disabled,
PXIe-5665 14 GHz VSA Preselector Disabled)
Center Frequency
As-Found Limit (dBm)
As-Left Limit (dBm)
*
>700 MHz to 3.6 GHz
+20
+21
>3.6 GHz to 14 GHz
†
+20
+21
*
Refer to the
section of this document for more information
about as-left limits.
†
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
Table 35.
TOI (Frequencies >700 MHz) Verification Test Limits (Preamplifier Enabled)
Center Frequency
As-Found Limit
As-Left Limit
*
>700 MHz to 3.6 GHz
+2.5 dBm
+3.5 dBm
*
Refer to the
section of this document for more information
about as-left limits.