36
|
ni.com
|
PXIe-5665 Calibration Procedure
Figure 12.
PXIe-5665 14 GHz VSA Average Noise Level Verification Equipment Setup
2.
Create a new session for the PXIe-5665.
3.
Configure the PXIe-5665 according to the following property settings:
•
Acquisition Type: Spectrum
•
Averaging Mode: RMS Averages
•
Channel Coupling: AC Coupled
•
Digital IF Equalization Enabled: TRUE
•
Digitizer Dither Enabled: Enabled
•
Number of Averages: 20
•
Ref Clock Source: PXI_Clk
•
Power Spectrum Units: Volts Squared
•
Preamp Enabled: Disabled
•
Resolution Bandwidth Type: ENBW
•
Reference Level: -50 dBm
•
Resolution Bandwidth: 300 Hz
•
RF Attenuation: 0 dB
•
FFT Window Type: Blackman-Harris
•
(PXIe-5665 14 GHz VSA)
Preselector Enabled: Disabled
1
50
Ω
termination
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD
SENSITIVE
OUT
ALL PORTS
50
Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD
SENSITIVE
TTL
6.3 Vp-p
MAX
2 Vp-p
NOM
+20 dBm MAX
50
Ω
50
Ω
50
Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD
SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50
Ω
+15 dBm MAX
25 V DC MAX
+15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC
+20 dBm MAX > 10 MHz
+10 dBm MAX < 10 MHz
K
K
1
J
J
J
J
J
J