24
|
ni.com
|
PXIe-5665 Calibration Procedure
3.
Connect power sensor B to the other output of the power splitter through the 3.5 mm (m)-
to-3.5 mm (m) adaptor, the 3.5 mm (f)-to-3.5 mm (f) adaptor, and the 20 dB attenuator. The
completed equipment setup is shown in Figure 7.
Figure 7.
RF Source Power (with Splitter and Attenuator) Characterization
Equipment Setup
4.
Set the RF source 2 power to 0 dBm.
1
RF Source 2
2
SMA (m)-to-SMA (m) Cable
3
Power Sensor A
4
Power Splitter
5
3.5 mm (m)-to-3.5 mm (m) Adaptor
6
20 dB Attenuator
7
3.5 mm (f)-to-3.5 mm (f) Adaptor
8
Power Sensor B
9
Power Meter
3
5
7
6
8
4
1
9
2