NI 5622 Calibration Procedure
Initial Setup
Refer to the
NI High-Speed Digitizers Getting Started Guide
for
information about how to install the software and hardware and how
to configure the device in MAX.
Self-Calibration
The NI 5622 includes precise internal circuits and references used during
self-calibration to adjust for time and temperature drift.
Note
Allow a 15 minute warm-up period before you begin self-calibration.
Note
Always self-calibrate the digitizer before you perform verification. NI-SCOPE
includes self-calibration example programs for LabVIEW, CVI, and Microsoft Visual C.
You can initiate self-calibration using the following methods:
•
MAX
•
NI-SCOPE Soft Front Panel (SFP)
•
NI-SCOPE
MAX
To initiate self-calibration from MAX, complete the following steps:
1.
Disconnect or disable any AC inputs to the digitizer.
2.
Launch MAX.
3.
Expand
My System»Devices and Interfaces»PXI System
.
4.
Select the device that you want to calibrate.
5.
Initiate self-calibration using one of the following methods:
•
Click
Self-Calibrate
in the upper right corner of MAX.
•
Right-click the name of the device in the MAX configuration tree
and select
Self-Calibrate
from the drop-down menu.
NI-SCOPE SFP
To initiate self-calibration from the NI-SCOPE SFP, complete the
following steps:
1.
Disconnect or disable any AC inputs to the digitizer.
2.
Launch the NI-SCOPE SFP, which is available at
Start»All Programs»
National Instruments»NI-SCOPE»SCOPE Soft Front Panel
.
3.
Select the device you want to calibrate using the Device Configuration
dialog box by selecting
Edit»Device Configuration
.