National Instruments Corporation
51
NI 5622 Calibration Procedure
Appendix A: Calibration Utilities
NI-SCOPE supports several calibration utilities you can use to retrieve
information about adjustments performed on the NI 5622, change the
external calibration password, and store small amounts of information in
the onboard EEPROM. Although you can retrieve some data using MAX,
you can retrieve all the data programmatically using NI-SCOPE functions.
MAX
To retrieve data using MAX, complete the following steps:
1.
Select the device from which you want to retrieve information from
My System»Devices and Interfaces»PXI System
.
2.
Select the
Calibration
tab in the lower right corner.
You should see information about the last date and temperature for both
external and self-calibration.
Table 9.
Input Parameters for Vertical Range External Adjustment
Iteration
Range (V)
Input Voltage (V
pk-pk
)
Input Frequency (MHz)
1
0.7
1
10
2
1
1.6
10
3
1.4
1.6
10
4
0.7
1
187
5
1
1.6
187
6
1.4
1.6
187
Table 10.
Input Parameters for Frequency Range External Adjustment
Iteration
Start Frequency
(MHz)
Stop Frequency
(MHz)
Frequency Step
(MHz)
1
0.9
224.9
1
2
225.9
815.9
10
3
119.9
307.4
0.5