NI 5622 Calibration Procedure
36
ni.com
7.
Configure the bandpass filter using the niScope Property Node.
8.
Commit all the parameter settings to hardware using the niScope
Commit VI.
9.
Configure the signal generator to generate an
Input Frequency
of a
1.3 V
pk-pk
sine wave for the current iteration in Table 7.
10. Wait the amount of time the manufacturer recommends for the output
of the signal generator to settle.
11. Initiate a waveform acquisition using the niScope Initiate
Acquisition VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_SetAttributeViBoolean
with
the following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
“0”
attributeId
:
NISCOPE_ATTR_BANDPASS_
FILTER_ENABLED
value
: The
Bandpass Filter
value listed in
Table 7 for the current iteration
LabVIEW VI
C/C++ Function Call
Call
niScope_Commit
with the following
parameter:
vi
: The instrument handle from
niScope_init
LabVIEW VI
C/C++ Function Call
Call
niScope_InitiateAcquisition
with the following parameter:
vi
: The instrument handle from
niScope_init