9.
Connect the 50 Ω SMB terminator (plug) to channel 4 of the PXIe-5172 and repeat steps
2 through 4 for each configuration listed in the
table.
10. Connect the 50 Ω SMB terminator (plug) to channel 5 of the PXIe-5172 and repeat steps
2 through 4 for each configuration listed in the
table.
11. Connect the 50 Ω SMB terminator (plug) to channel 6 of the PXIe-5172 and repeat steps
2 through 4 for each configuration listed in the
table.
12. Connect the 50 Ω SMB terminator (plug) to channel 7 of the PXIe-5172 and repeat steps
2 through 4 for each configuration listed in the
table.
Adjustment
This section describes the steps needed to adjust the PXIe-5172 to meet published
specifications.
Adjusting the PXIe-5172
1.
Call the niHsai Group A External Cal API v1 Host.lvclass:Open Ext Cal Session VI to
obtain an external calibration session. To perform an adjustment, you must specify the
calibration password. By default, the calibration password is
NI
.
2.
Complete the
Adjusting the 1 MΩ Compensation Attenuator
procedure on channel 0.
3.
Complete the
Adjusting the 1 MΩ DC Reference
4.
Complete the
procedure on channel 0.
5.
Complete the
procedure on channel 0. This procedure only needs to
run on one channel.
6.
Complete the
Adjusting the 1 MΩ Compensation Attenuator
procedure on channel 1.
7.
Complete the
Adjusting the 1 MΩ DC Reference
8.
Complete the
procedure on channel 1.
9.
Complete the
Adjusting the 1 MΩ Compensation Attenuator
procedure on channel 2.
10. Complete the
Adjusting the 1 MΩ DC Reference
11. Complete the
procedure on channel 2.
12. Complete the
Adjusting the 1 MΩ Compensation Attenuator
procedure on channel 3.
13. Complete the
Adjusting the 1 MΩ DC Reference
14. Complete the
procedure on channel 3.
Note
If you are adjusting the PXIe-5172 (8CH), proceed to the next step. If
you are adjusting the PXIe-5172 (4CH), proceed to step 27.
15. Complete the
Adjusting the 1 MΩ Compensation Attenuator
procedure on channel 4.
16. Complete the
Adjusting the 1 MΩ DC Reference
17. Complete the
procedure on channel 4.
18. Complete the
Adjusting the 1 MΩ Compensation Attenuator
procedure on channel 5.
19. Complete the
Adjusting the 1 MΩ DC Reference
20. Complete the
procedure on channel 5.
PXIe-5172 Calibration Procedure
|
© National Instruments
|
33