Table 1. PXIe-5164 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
BNC (f)-to-N
(f) adapter
Fairview
Microwave
SM3526
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
•
Frequency range: DC to
401 MHz
•
VSWR: <1.1
•
Impedance: 50 Ω
SMA (m)-to-
BNC (m)
adapter (x2)
Fairview
Microwave
SM4716
Test system
characterization
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
•
Frequency range: DC to
401 MHz
•
VSWR: <1.1
•
Impedance: 50 Ω
SMA feed-
thru
terminator
Pasternack PE6026
Adjustment
Verifications:
•
Passband
amplitude
flatness and
bandwidth
•
Amplitude: 10 dBm
•
Frequency: 50 kHz to
301 MHz
•
VSWR:
–
<1.1 at 100 MHz
–
<1.25 at 301 MHz
•
Impedance: 50 Ω
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-5164 meets
published specifications:
•
Allow a warm-up time of at least 15 minutes after the chassis is powered on. The warm-
up time ensures that the PXIe-5164 is at a stable operating temperature.
•
Allow all test instruments a warm-up time of at least the stated amount of time in their
specification document. The warm-up time ensures that the test instruments are at a stable
operating temperature.