Troubleshooting
If an issue persists after you complete a troubleshooting procedure, contact NI
technical support or visit
.
What Should I Do if the PXIe-4162 Doesn't Appear in MAX?
1. In the MAX configuration tree, expand
Devices and Interfaces
.
2. Expand the
Chassis
tree to see the list of installed hardware, and press <F5> to
refresh the list.
3. If the module is still not listed, power off the system, ensure that all hardware
is correctly installed, and restart the system.
4. Navigate to the Device Manager.
Operating System
Description
Windows 10/8.1
Right-click the Start button, and select
Device Manager
.
Windows 7
Select
Start
»
Control Panel
»
Device Manager
.
5. Verify the PXIe-4162 appears in the Device Manager.
1. Under an NI entry, confirm that a PXIe-4162 entry appears.
Note
If you are using a PC with a device for PXI remote control
system, under
System Devices
, also confirm that no error
conditions appear for the
PCI-to-PCI Bridge
.
2. If error conditions appear, reinstall NI-DCPower and the PXIe-4162.
Why Is the ACCESS LED Off When the Chassis Is On?
Before proceeding, verify that the PXIe-4162 appears in MAX.
If the ACCESS LED fails to light after you power on the chassis, a problem may exist
with the chassis power rails, a hardware module, or the LED.