Electrical

The following section provides information about the NI PXIe-1086 AC input.

Note

Refer to the 

NI PXIe-1086 User Manual 

for full specifications.

AC Input

Input voltage range ........................................... 100 to 240 VAC
Operating voltage range

1

.................................. 90 to 264 VAC

Input current rating ........................................... 12 to 6 A
Input frequency range....................................... 50/60 Hz
Operating frequency range

1

.............................. 47 to 63 Hz

Over-current protection .................................... All outputs protected from short circuit and 

overload with automatic recovery

Efficiency.......................................................... 70% typical

Power disconnect.............................................. The AC power cables provide main power 

disconnect. Do not position the equipment so 
that it is difficult to disconnect the power cord. 
The front panel power switch causes the internal 
chassis power supply to provide DC power to the 
CompactPCI/PXI Express backplane. You also 
can use the front panel terminal block 4-pin 
connector and power mode switch to control the 
internal chassis power supply.

Chassis Parts

The following figures show the key features of the NI PXIe-1086 chassis front and back panels.

1   

This operating range is guaranteed by design.

Содержание PXIe-1086

Страница 1: ...the Controller 7 Step 3 Install Modules 9 Step 4 Configure the System In Software Using MAX 10 Step 5 Connect Signals to Modules 10 Where To Go Next 10 Maintenance 10 Service Interval 10 Cleaning 11...

Страница 2: ...ircuits Operating personnel and service personnel must not remove protective covers when operating or servicing the chassis Adjustments and service to internal components must be undertaken by qualifi...

Страница 3: ...rom short circuit and overload with automatic recovery Efficiency 70 typical Power disconnect The AC power cables provide main power disconnect Do not position the equipment so that it is difficult to...

Страница 4: ...PXI Filler Panels Optional 9 Removable Feet 10 Fan Door Latch 11 PXI Express Hybrid Peripheral Slots 16x 12 PXI Express PXI Express System Timing Slot 13 PXI Express System Controller Slot 14 Ethernet...

Страница 5: ...ust be set to High to meet the maximum altitude specification Pollution Degree 2 For indoor use only Note Refer to the NI PXIe 1086 User Manual for full specifications Installing the PXI Express Syste...

Страница 6: ...ted spec the power switch LED blinks green Note If two system fans or both power supply fans fail the chassis shuts down automatically protecting the chassis and modules from damage due to overheating...

Страница 7: ...cord grounds the chassis and protects it from electrical damage while you install the controller Caution Refer to your PXI Express system controller user manual for specific instructions and warnings...

Страница 8: ...ler installed in the system controller slot of a chassis Figure 5 PXI Express System Controller Installed in an NI PXIe 1086 Chassis 4 Connect the keyboard mouse and monitor to the appropriate connect...

Страница 9: ...chassis making sure that the injector ejector handle is pushed down Figure 6 Installing PXI PXI Express or CompactPCI Peripheral Modules 5 When you begin to feel resistance push up on the injector ej...

Страница 10: ...abeled PXI System Unidentified Right click this entry to display the pop up menu then select the appropriate system controller model from the Identify As submenu 5 Click the PXI system controller The...

Страница 11: ...ts Use just enough moisture to dampen the cloth Caution Do not wash the front or rear panel connectors or switches Cover these components while cleaning the chassis Caution Do not use harsh chemical c...

Страница 12: ...Before handling the power supply allow the fan to stop spinning 1 Disengage the two captive screws on the rear of the power supply with a flat blade screwdriver 2 Extend the collapsible handle and pu...

Страница 13: ...remove the power drawer 1 Loosen the drawer lever captive screws with a flat blade screwdriver until the threads disengage from the chassis frame 2 Rotate the drawer levers to eject the drawer from th...

Страница 14: ...power drawer from the chassis The power drawer is shown in Figure 8 Note If you are using the NI PXIe 1086 and NI SC Express modules with front mounting terminal blocks together you must remove the SC...

Страница 15: ...our National Instruments product Product registration facilitates technical support and ensures that you receive important information updates from NI National Instruments corporate headquarters is lo...

Страница 16: ...s You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal expor...

Отзывы: