Appendix A: Calibration Utilities

NI-RFSG provides a full complement of external calibration VIs. You can use these VIs to 
retrieve information about adjustments performed on the PXI-5650/5651/5652. The External 
Calibration palette includes the following VIs:

niRFSG Initialize External Calibration VI

niRFSG Get External Calibration Last Date And Time VI

niRFSG Change External Calibration Password

niRFSG Update External Calibration Date and Time VI 

niRFSG Close External Calibration VI

Calibration VI References

The VIs used in this procedure, including all calibration VIs, are documented in the 

NI-RFSG VI 

Reference, 

which you can access from the 

NI RF Signal Generators Help

 at 

Start»

All Programs»National Instruments»NI-RFSG»Documentation

.

Worldwide Support and Services

The NI website is your complete resource for technical support. At 

ni.com/support

 you have 

access to everything from troubleshooting and application development self-help resources to 
email and phone assistance from NI Application Engineers.

Visit 

ni.com/services

 for NI Factory Installation Services, repairs, extended warranty, and 

other services.

Visit 

ni.com/register

 to register your NI product. Product registration facilitates technical 

support and ensures that you receive important information updates from NI.

A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European 
Communities using the manufacturer’s declaration of conformity. This system affords the user 
protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC 
for your product by visiting 

ni.com/certification

. If your product supports calibration, 

you can obtain the calibration certificate for your product at 

ni.com/calibration

.

NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 
78759-3504. NI also has offices located around the world. For telephone support in the United 
States, create your service request at 

ni.com/support

 or dial 1 866 ASK MYNI (275 6964). 

For telephone support outside the United States, visit the Worldwide Offices section of 

ni.com/niglobal

 to access the branch office websites, which provide up-to-date contact 

information, support phone numbers, email addresses, and current events.

Содержание PXI-5650

Страница 1: ...6 Verification 8 Validating the ALC Limits Table 8 Verifying Frequency Accuracy 9 Verifying High Power Level Accuracy 9 Verifying Low Power Level Accuracy 11 Verifying Single Sideband Phase Noise at...

Страница 2: ...e verification procedures Specifications for your device NI PXI 5650 5651 5652 Getting Started Guide NI RF Signal Generators Help The latest versions of these documents are available on ni com manuals...

Страница 3: ...with Preamplifier option B23 and B25 Characterizing the spectrum analyzer Frequency accuracy Low power level accuracy Single sideband phase noise Harmonics Subharmonics Nonharmonics Frequency range 50...

Страница 4: ...accuracy 1E 9 BNC m to BNC m cable Pomona 5697 SMA m to BNC m cable Pasterneck Enterprises PE3696 36 SMA m to SMA m cable MegaPhase G916 S1S1 36 Frequency range DC to 6 6 GHz Insertion loss 1 2 dB up...

Страница 5: ...time ensures that the PXI 5650 5651 5652 is at a stable operating temperature In each verification procedure insert a delay between configuring all devices and acquiring the measurement This delay may...

Страница 6: ...connectors on the device under test DUT and test equipment are fragile Carefully perform the steps in these procedures to prevent damaging any DUTs or test equipment Characterizing the Spectrum Analy...

Страница 7: ...frequency test points recorded in step 5 7 Connect the rubidium 10 MHz frequency reference signal to the spectrum analyzer REF IN rear panel connector 8 Connect the 6 dB precision attenuator to the PX...

Страница 8: ...on the device EEPROM 1 Call the niRFSG Initialize External Calibration VI 2 Call the niRFSG 5652 Validate ALC Limits Table VI Note Any invalid ALC limits are reported in the invalid limits indicator I...

Страница 9: ...ter setting the following NI RFSG property settings Frequency Hz 400 MHz Power Level dBm 20 dBm Generation Mode CW Reference Clock source Onboard Clock 6 Use the spectrum analyzer to measure the frequ...

Страница 10: ...High Power Level Accuracy Test Points and Limits Device Frequency Start Power dBm Test Limit dB PXI 5650 5651 5652 500 kHz 5 1 6 1 MHz 5 2 MHz 5 5 MHz 5 10 MHz 8 0 75 20 MHz 8 50 MHz 10 100 MHz 10 150...

Страница 11: ...sion attenuator to the PXI 5650 5651 5652 RF OUT front panel connector Note You must use the same attenuator you used in the Characterizing the Spectrum Analyzer procedure 3 Connect the spectrum analy...

Страница 12: ...r measurement in this step at the corresponding frequency Compare the value to the Test Limit listed in Table 5 Table 5 Low Level Accuracy Test Points and Limits 40 dBm Device Frequency Stop Power dBm...

Страница 13: ...PXI 5650 5651 5652 RF OUT front panel connector 3 Connect the spectrum analyzer RF INPUT connector to the 6 dB precision attenuator 4 Connect the rubidium 10 MHz frequency reference signal to the spe...

Страница 14: ...spectrum analyzer and record this value 10 Apply the following equation to the value measured in step 9 phase noiseSSB at 10 KHz offset Measurement 10log RBW where Measurement value measured in step...

Страница 15: ...1 Connect the PXI 5650 5651 5652 REF IN OUT front panel connector to the rubidium 10 MHz frequency reference signal 2 Connect the 6 dB precision attenuator to the PXI 5650 5651 5652 RF OUT front pane...

Страница 16: ...generated 12 Configure the spectrum analyzer with the following settings Center frequency 200 MHz Reference level 5 dBm Resolution bandwidth 500 Hz Span 100 kHz Reference Clock source External 13 Meas...

Страница 17: ...bration Procedure National Instruments 17 Table 8 Test Points for Nonharmonic Test Device Test Point MHz PXI 5650 5651 5652 200 400 600 800 1 000 1 200 PXI 5651 5652 1 400 1 600 1 800 2 000 2 200 2 40...

Страница 18: ...00 4 600 4 800 5 000 5 200 5 400 5 600 5 800 6 000 6 200 6 400 6 600 Table 9 Center Frequencies for Nonharmonic Test Device Center Frequency MHz PXI 5650 5651 5652 35 125 1 235 125 PXI 5651 5652 3 235...

Страница 19: ...ettings Center frequency 10 MHz Reference level 5 dBm Resolution bandwidth 1 kHz Frequency span 500 kHz Reference Clock source External 6 Generate a signal after setting the following NI RFSG property...

Страница 20: ...51 5652 10 MHz 15 20 MHz 50 MHz 100 MHz 150 MHz 200 MHz 400 MHz 800 MHz 1 0 GHz PXI 5651 5652 1 6 GHz 20 2 0 GHz 2 5 GHz 3 0 GHz 3 3 GHz PXI 5652 3 5 GHz 20 3 7 GHz 4 0 GHz 4 5 GHz 5 0 GHz 5 5 GHz 6 0...

Страница 21: ...th the following settings Center frequency 10 MHz Reference level 45 dBm Resolution bandwidth 50 Hz Frequency span 500 Hz Reference Clock source External 6 Generate a signal after setting the followin...

Страница 22: ...50 dBm to 0 dBm Output Power Device Frequency Test Limit dBc PXI 5650 5651 5652 10 MHz 15 20 MHz 50 MHz 100 MHz 150 MHz 200 MHz 400 MHz 800 MHz 1 0 GHz PXI 5651 5652 1 6 GHz 20 2 0 GHz 2 5 GHz 3 0 GH...

Страница 23: ...ear panel connector 5 Configure the spectrum analyzer with the following settings Center frequency 3 3 GHz Reference level 5 dBm Resolution bandwidth 1 kHz Span 500 kHz Reference Clock source External...

Страница 24: ...Test Limit dBc Subharmonic 1 5x Frequency GHz Subharmonic 1 5x Test Limit dBc 3 31 1 655 30 4 965 15 3 5 1 75 5 25 25 3 7 1 85 5 55 3 9 1 95 5 85 4 1 2 05 6 15 4 2 2 10 25 6 30 20 4 5 2 25 20 6 75 25...

Страница 25: ...ear panel connector 5 Configure the spectrum analyzer with the following settings Center frequency 3 3 GHz Reference level 45 dBm Resolution bandwidth 50 Hz Span 500 Hz Reference Clock source External...

Страница 26: ...x Test Limit dBc Subharmonic 1 5x Frequency GHz Subharmonic 1 5x Test Limit dBc 3 31 1 65 30 4 95 15 3 5 1 75 5 25 25 3 7 1 85 5 55 3 9 1 95 5 85 4 1 2 05 6 15 4 2 2 10 25 6 30 20 25 4 5 2 25 20 6 75...

Страница 27: ...able limits parameter to FALSE 4 Call the niRFSG 5652 Initialize ALC Calibration VI and set the measurement range parameter to Above 10 MHz 5 Call the niRFSG 5652 Configure ALC Calibration VI 6 Use th...

Страница 28: ...Calibration VI 6 Call the niRFSG 5652 Initialize Attenuator Calibration VI and set the measurement range parameter to Above 40 dBm Below 10 MHz 7 Call the niRFSG 5652 Configure Attenuator Calibration...

Страница 29: ...e corresponding frequency 23 Call the niRFSG 5652 Adjust Attenuator Calibration VI and wire the measured value from step 22 to the measured RF OUT power parameter 24 Repeat steps 20 through 23 until t...

Страница 30: ...urned by the center frequency parameter of the niRFSG 5652 Configure Reference Clock Calibration VI 9 Use the spectrum analyzer frequency counter to measure the frequency of the peak 10 Call the niRFS...

Страница 31: ...from NI Application Engineers Visit ni com services for NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your NI product Product registr...

Страница 32: ...an find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compl...

Отзывы: