34
|
ni.com
|
NI PXI-4130 Calibration Procedure
Figure 12.
Voltage Adjustment Connection Diagram
Table 18.
NI 4130 Output Parameters for Voltage Programming Accuracy Adjustment
Channel
Iteration
Voltage Level
Range (V)
Output
Current Limit
(A)
0
1
6
0 V
0.5
2
3 V
3
6 V
1
1
6
1 mV
2
3 V
3
6 V
4
-0.1 mV
5
-3 V
6
-6 V
1
20
1 mV
2
10 V
3
20 V
4
-1 mV
5
-10 V
6
-20 V
NI 41
3
0
+
–
LO
S
en
s
e
HI
S
en
s
e
HI
LO
DMM
Volt
a
ge
Mode