National Instruments Corporation
15
NI USB-9201/9221 User Guide and Specifications
Operating voltage ranges
1
NI USB-9201 .................................. ±10 V (not software selectable)
NI USB-9221 .................................. ± 60 V (not software selectable)
Maximum voltage, (AI or COM to earth ground, verified by a dielectric
withstand test)
Screw terminal ................................ 250 V
rms
, 5 seconds max
D-SUB ............................................ ±60 VDC
Isolation voltage (AI or COM to earth ground, verified by a dielectric
withstand test)
Screw terminal ................................ 2,300 V
rms
, 5 seconds max
D-SUB ............................................ 1,000 V
rms
, 5 seconds max
Overvoltage protection
(AI to COM)........................................... ±100 V
Accuracy
2
1
section for details about safe operating voltages.
2
Excludes noise.
Error
Percent of
Reading
(Gain Error)
Percent of
Range
(Offset
Error)*
NI USB-9201
Calibrated typ (25 °C, ±5 °C)
±0.04%
± 0.07%
Calibrated max (0 to 60 °C)
±0.25%
± 0.25%
Uncalibrated typ (25 °C, ±5 °C)
±0.26%
± 0.46%
Uncalibrated max (0 to 60 °C)
±0.67%
±1.25%
NI USB-9221
Calibrated typ (25 °C, ±5 °C)
±0.04%
±0.07%
Calibrated max (0 to 60 °C)
±0.25%
±0.25%
Uncalibrated typ (25 °C, ±5 °C)
±0.26%
±0.43%
Uncalibrated max (0 to 60 °C)
±0.67%
±1.06%