Chapter 1
Installing and Configuring the SCXI-1129
1-22
ni.com
already installed any other SCXI modules according to their installation
instructions. To install the SCXI-1129 module with an SCXI chassis,
follow these steps:
1.
Install the SCXI-1129s in the chassis and install the HVAB as
described in one of the Connecting the SCXI-1129 backplane sections.
2.
To connect an SCXI-1129 to an external DMM device, you need two
cables
—
the HV8-BAN4 or BAN2-WIRE cable, and a cable to connect
the handshaking lines from the SCXI-1129 to the DMM. Attach your
HV8-BAN4 cable to the HVAB connector on the HVAB-backplane
adapter, or connect the bare wire end of the BAN2-WIRE cable to one
or more rows on one of the terminal blocks.
3.
Attach the other end of the HV8-BAN4 or BAN2-WIRE cable to the
inputs of your external DMM device.
4.
Connect the handshaking lines from the front of either the SCXI-1333,
SCXI-1334, SCXI-1335, SCXI-1336, SCXI-1337, or SCXI-1339
terminal block to the appropriate lines on the external DMM.
5.
Check the installation.
6.
Power on the SCXI chassis.
7.
Power on the computer or reconnect it to your chassis.
The SCXI-1129 module is now installed.
Note
For the system to function properly, it must be controlled by a DAQ device or
NI 4021 switch controller, or be a SCXI-2000 chassis.
Configuring and Self-Test
Launch MAX to configure and test your SCXI-1129. If you need help
during the configuration process, open the Measurement & Automation
Help file by selecting
Help
»
Help Topics
.
1.
Double-click the
Measurement & Automation Explorer
icon on
your desktop.
2.
To add a new chassis, right-click
Devices and Interfaces
and select
Create New
.
3.
Select the appropriate chassis from the list box and click
Finish
.