NI PXIe-4464 Calibration Procedure
|
© National Instruments
|
3
Table 1.
Recommended Equipment
Equipment
Recommended
Model(s)
Where Used
Minimum
Requirements
Calibrator
Fluke 5700A
AI Gain
Frequency Range:
40 Hz to 92.2 kHz
Voltage Range:
up to 6.3 V
rms
ACV Accuracy:
40 Hz: 0.017%
1 kHz: 0.016%
20 kHz: 0.017%
45 kHz: 0.087%
92.2 kHz: 0.29%
Function Generator
Keysight (Agilent)
33250A series
Timebase
Frequency
Frequency Range:
up to 90 kHz
Frequency Accuracy:
2 ppm
Voltage Range:
up to 9 V
pk
Digital Multimeter
NI PXI-4070
IEPE Current
Current Accuracy: 0.9%
BNC Shorting Cap
(quantity 4)
Pomona Electronics
5085
or Pomona
Electronics 3840-50
AI Offset
Resistance:
≤
50
Ω
BNC F-F-F T-Connector
(quantity 3)
Pomona Electronics
3284
AI Gain
Characteristic
Impedance: 50
Ω
BNC (F) to Banana Adapter
(quantity 1)
Pomona Electronics
1269
AI Gain
—
BNC (M) Cable
(BNC variant: quantity 7
mXLR variant: quantity 3)
Pomona Electronics
5697
AI Gain
Characteristic
Impedance: 50
Ω
mXLR (F) to BNC (M) Cable
*
(quantity 4)
NI 156789-XX or
NI 140150-XX
0R460.46 m
0R910.91 m
2R42.4 m
All
—
BNC F-F Adapter
*
(quantity 4)
Pomona Electronics
3283
AI Offset
—
*
Required only for the mXLR variant.