22
|
ni.com
|
NI 5122/5124/5142 Calibration Procedure
17. Move the calibrator test head to the digitizer input channel 1 and repeat steps 2 through 16,
changing the value of the
channelList
parameter from
"0"
to
"1"
.
18. End the session using the niScope Close VI.
You have finished verifying the programmable vertical offset accuracy of the
NI 5122/5124/5142.
Timing Accuracy
Complete the following steps to verify the timing accuracy for the NI 5122/5124/5142.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
2.
Configure the input impedance and the maximum input frequency using the
niScope Configure Chan Characteristics VI.
Table 10.
NI 5124 Programmable Vertical Offset Accuracy Limits
Calibration Test Limits
Published Specifications
±0.36%
±0.4%
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the
following parameter:
vi
: The instrument handle from
niScope_init
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the
following parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE