Vertical Gain Accuracy
Table 6 contains the input parameters for verifying vertical gain accuracy, the calibration test
limits, and the published specifications of the NI 5105.
To verify vertical gain accuracy, complete the procedures described in this section for each of
the 12 iterations listed in Table 6 for channel 0, then repeat for each of the remaining 7 channels
on the NI 5105.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
2.
Configure the input impedance and the maximum input frequency using the
niScope Configure Chan Characteristics VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the
following parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the
following parameters:
channelList
: The name of the
channel being tested
inputImpedance
: The
Input
Impedance
value listed in
Table 6 for the current iteration
maxInputFrequency
: The
Max Input Frequency
value
listed in Table 6 for the current
iteration