![National Instruments NI PXI-4220 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/national-instruments/ni-pxi-4220/ni-pxi-4220_user-manual_709323060.webp)
Chapter 4
Theory of Operation
4-22
ni.com
For more detailed information about the operation of any of these circuitry
stages, refer to the
Bridge Configuration and Completion
,
Excitation
,
,
and
sections.
Bridge Configuration and Completion
You can configure the NI PXI-4220 for use with Wheatstone bridge
sensors that require bridge completion. Bridge completion is necessary for
quarter- or half-bridge sensors. You can set the NI PXI-4220 for quarter-,
half-, or full-bridge configuration to match the configuration completion
requirements of each sensor. When quarter- or half-bridge configuration is
selected, S
X–
(pin 6) (where
X
is a particular channel) is disconnected from
the front signal connector and internally connected to a half-bridge
completion network. When quarter-bridge configuration type I is selected,
a socketed quarter-bridge completion resistor is internally connected
between P
X
– (pin 7) and QTR/SCB
X
(pin 9). You then field wire the
quarter-bridge sensor between S
X
+ (pin 1) and P
X
+ (pin 2) with a third
lead connected to QTR/SCB
X
(pin 9) as shown in Figure 2-3,
Quarter-Bridge I Circuit Diagram
. Make sure that the value of the
precision quarter-bridge completion resistor exactly matches the nominal
gauge resistance of the strain gauge. The quarter-bridge completion resistor
is socketed for easy replacement. The quarter-bridge completion resistor
for AI 0 is located on the NI PXI-4220 at reference designator
R5
, the
completion resistor for AI 1 is located at
R2
.
Note
The NI PXI-4220 ships with a 350
Ω
quarter-bridge completion resistor installed.
Refer to the
section of Chapter 3,
, for more information about programmatically
setting bridge completion in MAX. For more information about
programmatically setting bridge completion in NI-DAQmx, refer to the
Developing Your Application in NI-DAQmx
section of Chapter 5,
.
Excitation
The NI PXI-4220 provides DC excitation voltage for a Wheatstone bridge
sensor. For half- and full-bridge applications, the excitation voltage is
available at pins P
X
+ (pin 2) and P
X
– (pin 7). For quarter-bridge
applications, P
X
– (pin 7) is not used; instead, wire to S
X
+ (pin 1)
and P
X
+ (pin 2) and short S
X
+ (pin 1) to pin QTR/SCB
X
(pin 9).
Pin QTR/SCB
X
(pin 9) internally connects to P
X