1-28
|
ni.com
Chapter 1
Getting Started with the cDAQ Chassis
4.
Power on the MXI-Express device.
5.
Power on the cDAQ chassis.
The MXI-Express port has one LINK LED. Refer to Table 1-8 for information about the
MXI-Express LINK LED behavior.
Note
Do
not
connect MXI-Express devices to the cDAQ chassis while the chassis
is powered on. The MXI-Express device may not be detected by the cDAQ chassis.
DIP Switches
The cDAQ chassis features six DIP switches, shown in Figure 1-1.
Table 1-7.
DIP Switches
Switch
Description
DISABLE
RT
(NI cDAQ-9138/9139 for LabVIEW Real-Time)
The position of the
DISABLE RT determines the operating system the cDAQ chassis boots
into. If the switch is in the OFF position, the chassis boots into LabVIEW
RT. Move the switch to the ON position to boot the cDAQ chassis from an
external drive.
(NI cDAQ-9138/9139 for Windows)
The position of the DISABLE RT
determines the operating system the cDAQ chassis boots into. If the switch
is the ON position, the chassis boots into Windows Embedded Standard 7
(WES7).
If the switch is in the OFF position, the chassis does not boot into Windows.