Physical Characteristics
Dimensions
and search by module number.
Weight
Approx. 144 g (5.0 oz)
Safety Voltages
Connect only the voltages that are within these limits:
Supply voltage range (Vsup)
+8 V DC to +18 V DC
Maximum voltage
Port-to-COM
-40 V DC to +40 V DC maximum, Measurement Category I
Isolation
Port-to-earth ground
Withstand
1,000 V RMS, verified by a 5 s dielectric withstand test
Continuous
60 V DC, Measurement Category I
Safety Compliance and Hazardous Locations Standards
This product is designed to meet the requirements of the following electrical
equipment safety standards for measurement, control, and laboratory use:
■
IEC 61010-1, EN 61010-1
■
UL 61010-1, CSA C22.2 No. 61010-1
■
EN 60079-0, EN 60079-7
■
IEC 60079-0, IEC 60079-7
■
UL 60079-0, UL 60079-7
■
CSA C22.2 No. 60079-0, CSA C22.2 No. 60079-7