b)

Configure the task timing according to the following 

NI 9251 with mini XLR Timing

Configuration

 table.

Table 3. NI 9251 with mini XLR Timing Configuration

Sample Mode

Samples Per Channel

Rate (kS/s)

 

 

 

Finite Samples

10240

102.4

c)

Start the task.

d)

Read the samples from the NI 9251 with mini XLR.

e)

Average the samples for each AI channel.

f)

Clear the task.

6.

Set the calibrator to Standby mode (STBY).

7.

Compare the averaged samples to the limits in the 

NI 9251 with mini XLR Verification

Test Limits

 table to verify the input accuracy. If the value is within these limits, the device

passes the test.

8.

Repeat steps 3 through 7 for each test point in the 

NI 9251 with mini XLR Verification

Test Limits

 table.

9.

Disconnect the calibrator from the device.

Table 4. NI 9251 with mini XLR Verification Test Limits

Test Point

2-Year Limits

Location

Value (V)

Lower Limit (VDC)

Upper Limit (VDC)

 

 

 

 

Minimum

-1.9

-1.905161

-1.894839

Mid-scale

0

-0.002885

0.002885

Maximum

1.9

1.894839

1.905161

Note

 

The test limits listed in the 

NI 9251 with mini XLR Verification Test Limits

table are derived using values from the 

NI 9251 with mini XLR Accuracy Under

Calibration Conditions

 table.

Adjustment

The following performance adjustment procedure describes the sequence of operation required
to adjust the NI 9251 with mini XLR.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the
NI 9251 with mini XLR.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9251 with mini XLR to the calibrator as shown in the 

Accuracy

Verification

 section.

NI 9251 Calibration Procedure 

  | 

© National Instruments

  | 

5

Содержание NI 9251

Страница 1: ...e 6 Reverification 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9251 with mini XLR requires the installation of NI DAQmx 16 1 or later on the calibration system You can download NI DAQmx from ni com downloads NI DAQmx supports LabVIEW LabWindows CVI ANSI C and NET When you install NI DAQmx you only need to install support for the applicatio...

Страница 2: ...nd NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test Equipment The following table lists the equipment recommended for the performance verification and adjustment procedures If the recommended equipment is not available select a substitute using the re...

Страница 3: ...copper wire for all cable connections to the device Use twisted pairs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C The device temperature will be greater than the ambient temperature Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the device measurement circuitry is at a stable operating temperature Initial Setup...

Страница 4: ...mine the As Found status of the NI 9251 with mini XLR 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9251 with mini XLR to the calibrator as shown in the following figure Figure 1 Analog Input Verification Connections NI 9251 Calibrator AI0 AI0 Chassis GND HI LO AI1 AI1 Note The chassis ground lug connection to ground is the NI 9251 with mini XLR ground connection 3 Set the calibrator ...

Страница 5: ...Disconnect the calibrator from the device Table 4 NI 9251 with mini XLR Verification Test Limits Test Point 2 Year Limits Location Value V Lower Limit VDC Upper Limit VDC Minimum 1 9 1 905161 1 894839 Mid scale 0 0 002885 0 002885 Maximum 1 9 1 894839 1 905161 Note The test limits listed in the NI 9251 with mini XLR Verification Test Limits table are derived using values from the NI 9251 with mini...

Страница 6: ...point in the array i Close the calibration session 4 Disconnect the NI 9251 with mini XLR from the calibrator EEPROM Update When an adjustment procedure is completed the NI 9251 with mini XLR internal calibration memory EEPROM is immediately updated If you do not want to perform an adjustment you can update the calibration date and onboard calibration temperature without making any adjustments by ...

Страница 7: ...t Product registration facilitates technical support and ensures that you receive important information updates from NI A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC for y...

Страница 8: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to obtain relevant HTS codes ECCNs and other import export data NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY...

Отзывы: