Caution
Do not mix hazardous voltage circuits and human-accessible
circuits on the same module.
Caution
Ensure that devices and circuits connected to the module are
properly insulated from human contact.
Caution
When module terminals are hazardous voltage LIVE (>42.4 V
pk
/60
V DC), you must ensure that devices and circuits connected to the module
are properly insulated from human contact.
Safety Voltages
Connect only voltages that are within the following limits:
Maximum working voltage, channel-to-earth ground
Continuous
Up to 2,000 m altitude 300 Vrms, Measurement Category III
Up to 5,000 m altitude 150 Vrms, Measurement Category III or 300 Vrms, Measurement
Category II
Maximum working voltage, channel-to-channel
Continuous
Up to 2,000 m altitude 480 Vrms, Measurement Category III
Up to 5,000 m altitude 300 Vrms, Measurement Category III or 480 Vrms, Measurement
Category II
Withstand
Channel-to-channel
3510 Vrms, verified by a 5 s dielectric withstand test
Channel-to-earth ground
3510 Vrms, verified by a 5 s dielectric withstand test