background image

Caution

 The system shall only be used in an area of not more than

Pollution Degree 2, as defined in IEC/EN 60664-1.

Caution

 The system shall be mounted in an ATEX/IECEx-certified

enclosure with a minimum ingress protection rating of at least IP54 as

defined in IEC/EN 60079-15.

Caution

 The enclosure must have a door or cover accessible only by the

use of a tool.

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits

for electromagnetic compatibility (EMC) stated in the product specifications. These

requirements and limits provide reasonable protection against harmful interference

when the product is operated in the intended operational electromagnetic

environment.
This product is intended for use in industrial locations. However, harmful

interference may occur in some installations, when the product is connected to a

peripheral device or test object, or if the product is used in residential or

commercial areas. To minimize interference with radio and television reception and

prevent unacceptable performance degradation, install and use this product in strict

accordance with the instructions in the product documentation.
Furthermore, any changes or modifications to the product not expressly approved

by National Instruments could void your authority to operate it under your local

regulatory rules.

Special Conditions for Marine Applications

Some products are approved for marine (shipboard) applications. To verify marine

approval certification for a product, visit 

ni.com/product-certifications

, search by

model number, and click the appropriate link.

ni.com

6

NI-9220 Getting Started

Содержание NI 9220

Страница 1: ...NI 9220 Getting Started 2022 07 06 ...

Страница 2: ...cial Conditions for Marine Applications 6 Preparing the Environment 7 NI 9220 Pinout 7 Grounded Differential Connections 8 Floating Differential Connections 8 Single Ended Connections 8 NI 9220 Connection Guidelines 8 Connecting to a Spring Terminal Connector 8 High Vibration Application Connections 9 Overvoltage Protection 9 NI Services 9 ni com 2 NI 9220 Getting Started ...

Страница 3: ...tion Observe all instructions and cautions in the user documentation Using the product in a manner not specified can damage the product and compromise the built in safety protection Attention Suivez toutes les instructions et respectez toutes les mises en garde de la documentation d utilisation L utilisation du produit de toute autre façon que celle spécifiée risque de l endommager et de compromet...

Страница 4: ...sure that devices and circuits connected to the product are properly insulated from human contact Attention Lorsqu une haute tension dangereuse est appliquée aux bornes du produit vous devez vous assurer que les appareils et les circuits auxquels il est connecté sont correctement isolés de tout contact humain Caution You must use the connector backshell kit to ensure that the terminals are not acc...

Страница 5: ...s connected signals must be within the following limits Capacitance 0 2 µF maximum Special Conditions for Hazardous Locations Use in Europe and Internationally The NI 9220 has been evaluated as equipment under DEMKO ATEX and is IECEx certified Each NI 9220 is marked and is suitable for use in Zone 2 hazardous locations in ambient temperatures of 40 C Ta 70 C If you are using the NI 9220 in Gas Gro...

Страница 6: ...ironment This product is intended for use in industrial locations However harmful interference may occur in some installations when the product is connected to a peripheral device or test object or if the product is used in residential or commercial areas To minimize interference with radio and television reception and prevent unacceptable performance degradation install and use this product in st...

Страница 7: ...e desired EMC performance is attained Preparing the Environment Ensure that the environment in which you are using the NI 9220 meets the following specifications Operating temperature IEC 60068 2 1 IEC 60068 2 2 40 C to 70 C Operating humidity IEC 60068 2 78 10 RH to 90 RH noncondensing Pollution Degree 2 Maximum altitude 2 000 m Indoor use only Note Refer to the device datasheet on ni com manuals...

Страница 8: ...I14 COM AI3 AI5 AI7 AI9 AI11 AI13 AI15 COM 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 Signal Description AI Positive analog input signal connection AI Negative analog input signal connection COM Common reference connection to isolated ground Table 1 Signal Descriptions Grounded Differential Connections Floating Differential Connections Single...

Страница 9: ... Remove the screwdriver from the activation slot to clamp the wire into place High Vibration Application Connections If your application is subject to high vibration NI recommends that you use the backshell kit to protect connections to the NI 9220 with spring terminal Overvoltage Protection The NI 9220 provides overvoltage protection for each channel Note Refer to the NI 9220 Specifications on ni...

Страница 10: ...i com register to register your NI product Product registration facilitates technical support and ensures that you receive important information updates from NI NI corporate headquarters is located at 11500 N Mopac Expwy Austin TX 78759 3504 USA ni com 10 NI 9220 Getting Started 2022 National Instruments Corporation ...