The shield grounding methodology can vary depending on the application.

Refer to your thermocouple documentation or the thermocouple wire spool

to determine which wire is the positive lead and which wire is the negative

lead.

Minimizing Thermal Gradients

Changes in the ambient air temperature near the front connector or a thermocouple

wire conducting heat directly to terminal junctions can cause thermal gradients.

Observe the following guidelines to minimize thermal gradients and improve the

system accuracy.

Use small-gauge thermocouple wire. Smaller wire transfers less heat to or

from the terminal junction.

Run thermocouple wiring together near the TB-9214 to keep the wires at the

same temperature.

Avoid running thermocouple wires near hot or cold objects.

Minimize adjacent heat sources and air flow across the terminals.

Keep the ambient temperature as stable as possible.

Make sure the NI 9214 terminals are facing forward or upward.

Keep the NI 9214 in a stable and consistent orientation.

Allow the thermal gradients to settle after a change in system power or in

ambient temperature. A change in system power can happen when the

system powers on, the system comes out of sleep mode, or you insert/remove

modules.

If possible, use the foam pad in the TB-9214 opening to restrict airflow

around the terminals.

Wiring the TB-9214

What to Use

TB-9214

0.05 mm

2

 to 0.5 mm

2

 (30 AWG to 20 AWG) wire with 5.1 mm (0.2 in.) of the

inner insulation stripped and 51 mm (2.0 in.) of the outer insulation stripped

© National Instruments

9

NI-9214

Содержание NI 9214

Страница 1: ...NI 9214 2023 04 25...

Страница 2: ...5 Electromagnetic Compatibility Guidelines 6 Special Conditions for Marine Applications 7 Preparing the Environment 7 TB 9214 Pinout 7 Thermocouple Connections 8 NI 9214 Connection Guidelines 8 Minimi...

Страница 3: ...entire system 2010 2016 National Instruments Corporation All rights reserved Refer to the National Instruments _Legal Information directory for information about NI copyright patents trademarks warra...

Страница 4: ...circuits and human accessible circuits on the same module Caution Ensure that devices and circuits connected to the module are properly insulated from human contact Caution When module terminals are h...

Страница 5: ...nd Ex nA IIC T4 hazardous locations and nonhazardous locations only Follow these guidelines if you are installing the NI 9214 in a potentially explosive environment Not following these guidelines may...

Страница 6: ...than Pollution Degree 2 as defined in IEC 60664 1 Caution The system shall be mounted in an ATEX IECEx certified enclosure with a minimum ingress protection rating of at least IP54 as defined in IEC...

Страница 7: ...for the LR certificate or look for the Lloyd s Register mark on the product Caution In order to meet the EMC requirements for marine applications install the product in a shielded enclosure with shie...

Страница 8: ...1 Signal Descriptions Signal Description COM Common reference connection TC Positive thermocouple connection TC Negative thermocouple connection Thermocouple Connections Thermocouple TC COM TC Shield...

Страница 9: ...to keep the wires at the same temperature Avoid running thermocouple wires near hot or cold objects Minimize adjacent heat sources and air flow across the terminals Keep the ambient temperature as sta...

Страница 10: ...en the screw for the terminal Make sure no exposed wire extends past the screw terminal 3 Route the wire through the TB 9214 opening remove slack from the wiring and secure the wires using a zip tie 4...

Страница 11: ...f the European Communities using the manufacturer s declaration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC fo...

Отзывы: