National Instruments NI 6703 Скачать руководство пользователя страница 5

©

 National Instruments Corporation

5

NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

2.

Adjust the NI 6703/6704 calibration constants with respect to a 
known voltage source. This step is described in the 

Adjusting the 

NI 6703/6704

 section.

3.

Reverify the performance to ensure that the NI 6703/6704 is operating 
within its specifications after adjustment.

Verifying the Performance of the NI 6703/6704

Verification determines how well the device is meeting its specifications. 
The verification procedure is divided into the major functions of the device. 
Throughout the verification process, refer to the tables in the 

Specifications

 

section to see if the device needs to be adjusted.

Verifying Voltage Output

This procedure verifies whether the output of voltage channel 0 is within 
specifications. Because of the channel-to-channel matching of the 
NI 6703/6704 voltage channels, if voltage channel 0 is operating within 
specifications, all the voltage channels should be operating within 
specifications. 

If you want to verify the channels individually, this procedure can be easily 
expanded to verify all the voltage channels. Perform the procedure once for 
each value of the channel parameter from 0 to 15. Each time you repeat the 
procedure, you also need to connect the appropriate channel to the DMM. 
Refer to the

 PCI/PXI-6703/6704 User Manual

 at 

ni.com/manuals

 for the 

pin numbers corresponding to each channel. To verify the voltage output, 
read the 

Equipment and Other Test Requirements

 section, and complete the 

following steps:

1.

Call the 

Get_Cal_Date

 function contained in 

ni6704Cal.dll

 to 

determine the previous calibration date of the NI 6703/6704.

2.

Connect the DMM to the NI 6703/6704 as shown in the following 
table.

3.

Place the DMM in voltage mode.

4.

Call 

AO_VWrite

 to generate 9.9 V on voltage channel 0. Set the 

following parameters:

chan

: 0

voltage

: 9.9

Signal Name

Description

NI 6703/6704 Output

DMM Input

VCH0

Voltage Channel 0

Pin 34

High

AGND0

Analog Ground 0

Pin 68

Low

Содержание NI 6703

Страница 1: ...6704 products have two calibration options an internal or self calibration and an external calibration Internal Calibration The NI 6703 6704 is continuously applying corrections to the output levels b...

Страница 2: ...e NI recommends that you use the Hewlett Packard 3458A DMM for calibration If you do not have a Hewlett Packard 3458A DMM use the accuracy specifications to select a substitute calibration standard If...

Страница 3: ...constants update the calibration date and write to the factory calibration area The factory calibration area and the calibration date should only be modified by a metrology laboratory or another faci...

Страница 4: ...ode If you do not want to use these variable definitions you can examine the function call listings in the NI DAQ documentation and the nidaqcns h file to determine what input values are required Docu...

Страница 5: ...e channel 0 is operating within specifications all the voltage channels should be operating within specifications If you want to verify the channels individually this procedure can be easily expanded...

Страница 6: ...from the NI 6703 6704 You have now verified the voltage output of the NI 6703 6704 Verifying Current Output NI 6704 Only This procedure verifies that the current output of channel 16 is within specif...

Страница 7: ...rom the NI 6704 You have now verified the current output of the NI 6704 Adjusting the NI 6703 6704 At the end of each calibration procedure new constants are stored in the user calibration area of the...

Страница 8: ...age channel 0 Decreasing the voltage value increases the magnitude of the voltage on voltage channel 0 6 Call AO_VWrite to generate 10 0 V on voltage channel 0 Set the following parameters chan 0 volt...

Страница 9: ...Call AO_VWrite to generate 0 1 mA on current channel 16 Set the following parameters chan 16 voltage 0 1 5 Call AO_VWrite with the channel set to the current offset channel Adjust the voltage paramet...

Страница 10: ...values should also be stored in the factory calibration area Calibration constants stored in the factory calibration area cannot be overwritten by the user To save the calibration constants and updat...

Страница 11: ...mn contains the upper limits and lower limits for the test point value If the NI 6703 6704 has been calibrated in the last 24 hours the test point value should be between the upper and lower limit val...

Страница 12: ...ructure Voltage Output Verification Figure 1 Voltage Output Verification Table 2 Current Output Specifications NI 6704 Only Test Point Value mA 24 Hour Ranges 1 Year Ranges Lower Limit mA Upper Limit...

Страница 13: ...6704 Only Figure 2 Current Output Verification NI 6704 Only Refer to Table 2 Current Output Specifications NI 6704 Only to determine the appropriate ranges for the voltages and currents to verify Con...

Страница 14: ...voltage channel 0 AO_VWrite deviceNumber 0 0 0 NI 6703 AO_VWrite deviceNumber 16 voltage Adjust voltage until the DMM reads 0 0 V NI 6704 AO_VWrite deviceNumber 32 voltage Adjust voltage until the DMM...

Страница 15: ...Output Adjustment NI 6704 Only AO_VWrite deviceNumber 35 voltage Adjust voltage until the DMM reads 20 0 mA Set_DAQ_Device_Info deviceNumber ND_CALIBRATION_ENABLE ND_YES Connect the multimeter to curr...

Страница 16: ...ents covering National Instruments products refer to the appropriate location Help Patents in your software the patents txt file on your CD or ni com patents 2002 2004 National Instruments Corp All ri...