NI 6601/6602 Calibration Procedure
8
ni.com
e.
Program counter 1 to arm it.
Setting Up an Alternative Clock Source
To set up a clock source other than a PXI-6608, complete the following
steps:
1.
Ensure that the clock source meets the requirements listed in the
section.
2.
Power the clock device.
Note
After you complete the clock setup procedure, the clock is running and generating
pulses.
Connecting the Clock Source to the NI 6601/6602 Device
This section describes how to connect the clock source to the NI 6601/6602
device, whether you are using a PXI-6608 or another clock source.
Connecting a PXI-6608 to the NI 6601/6602 Device
Connect the PXI-6608 clock source to the NI 6601/6602 device using an
SH68-68-D1 cable, which connects to the I/O connector of each device.
Traditional NI-DAQ Function Call
LabVIEW Block Diagram
Call
GPCTR_Control
with the following
parameters:
deviceNumber
: The value set by MAX
gpctrNum
:
ND_COUNTER_1
action
:
ND_PROGRAM
“TASK ID” out connects to “TASK ID” in of step 14
in the