NI 5761R User Guide and Specifications
6
ni.com
The NI 5761 ships with socketed CLIP items that are used to add module I/O to the LabVIEW project.
The NI 5761 ships with the following CLIP items:
•
NI 5761 Multiple Sample CLIP
—Generates two samples per clock cycle at a clock rate that is
half the sample rate. The default sample rate is 250 MHz, which sets the default clock rate for this
CLIP at 125 MHz. You can set a lower sample rate by using an external Sample clock. This CLIP
provides access to four analog input channels, eight PFI lines, and an input clock selector that can
be configured to use one of the following settings:
–
Internal Sample clock
–
Internal Sample clock locked to an external Reference clock through the CLK IN connector
–
External Sample clock through the CLK IN connector
–
Internal Sample clock locked to an external Reference clock through IoModSyncClock
–
External Sample clock through IoModSyncClock
This CLIP also contains an engine to program the CLK chip and ADCs, either through predetermined
settings for an easier instrument setup or through a raw SPI address and data signals for a more advanced
setup. The NI 5761 Multiple Sample CLIP is the default CLIP.
•
NI 5761 Single Sample CLIP
—Generates one sample per clock cycle at a default sample rate of
250 MHz. You can set a lower sample rate by using an external Sample clock. This CLIP provides
access to four analog input channels, eight PFI lines, and an input clock selector that can be
configured to use one of the following settings:
–
Internal Sample clock
–
Internal Sample clock locked to an external Reference clock through the CLK IN connector
–
External Sample clock through the CLK IN connector
–
Internal Sample clock locked to an external Reference clock through IoModSyncClock
–
External Sample clock through IoModSyncClock
This CLIP also contains an engine to program the CLK chip and ADCs, either through predetermined
settings for an easier instrument setup or through a raw SPI address and data signals for a more advanced
setup.
Refer to the
NI FlexRIO Help
in the
LabVIEW Help
for more information about NI FlexRIO CLIP items,
configuring the NI 5761 with a socketed CLIP, and a list of available socketed CLIP and provided
signals.
Cables
Use the NI CMD-to-CMD cable (NI part number 763194-01) or the NI CMD Single-ended to Pigtail
cable (NI part number 763191-01) to connect signals to your AUX I/O connector. Use any 50
Ω
SMA
cable to connect signals to the other connectors on the front panel of your NI 5761. For more information
about connecting I/O signals on your device, refer to the
section of this document.