NI 5422 Specifications
24
ni.com
Start Trigger
DDC CLK IN
Clocking
Format
DDC Data Output signals change on the rising edge of
DDC CLK IN.
—
Frequency
Range
10 Hz to 200 MHz
—
Input Duty
Cycle
Tolerance
40% to 60%
—
Table 8.
Specification
Value
Comments
Sources
1. PFI<0..1> (SMB front panel connectors)
2. PFI<2..3> (DIGITAL DATA & CONTROL front panel
connector)
3. PXI_Trig<0..7> (backplane connector)
4. PXI Star trigger (backplane connector)
5. Software (use function call)
6. Immediate (does not wait for a trigger). Default.
—
Modes
1. Single
2. Continuous
3. Stepped
4. Burst
—
Edge Detection
Rising
—
Minimum
Pulse Width
25 ns. Refer to t
s1
at
NI Signal Generators Help»Devices»
NI 5422»NI PXI-5422»Triggering»Trigger Timing
.
—
Table 7.
(Continued)
Specification
Value
Comments