![National Instruments NI 447 Series Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/national-instruments/ni-447-series/ni-447-series_user-manual_3489585058.webp)
Appendix A
Specifications
A-8
ni.com
Physical
Dimensions (not including connectors)
NI PCI-4472/4474 ...........................17.5 by 10.7 cm (6.9 by 4.2 in.)
NI PXI-4472/4472B ........................16.0 by 9.9 cm (6.3 by 3.9 in.)
(1 3U CompactPCI slot)
Analog I/O connectors............................SMB male
Digital trigger connector.........................SMB male
Calibration
Internal....................................................On software command; computes
gain and offset corrections
Interval.............................................Whenever temperature is
different from temperature at
last internal calibration by more
than ±5 °C
External...................................................Internal voltage reference read
and stored in non-volatile
memory
Interval.............................................1 year
Warm-up time.........................................15 minutes
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth .....................................42 V, Installation Category I
Channel-to-channel.................................42 V, Installation Category I
Environmental
Operating temperature ............................0 to 50 °C
Storage temperature ................................–20 to 70 °C
Humidity .................................................10 to 90% RH, noncondensing