Chapter 2
Using the NI 447X
©
National Instruments Corporation
2-3
Caution
The NI PXI-4472 has connections to several reserved lines on the CompactPCI
J2 connector. Use of these lines by other devices in the CompactPCI system can damage
the equipment. Before installing the NI PXI-4472 in a CompactPCI system that uses
J2 connector lines for purposes other than PXI, refer to the
section of Chapter 1,
Getting Started with the NI 447X
.
3.
Make sure there are no lighted LEDs on the chassis. If any are lit, wait
until they go out before continuing your installation.
4.
Remove the filler panel for the slot you have chosen.
5.
Ground yourself using a grounding strap or by touching a grounded
object. Follow the ESD protection precautions described in the
Getting Started with the NI 447X
.
6.
Insert the NI PXI-4472 into a 5 V PXI slot. Use the injector/ejector
handle to fully insert the device into the chassis.
7.
Screw the front panel of the NI PXI-4472 to the front panel-mounting
rail of the system.
Note
To ensure a good ground connection, securely fasten the front panel of the
NI PXI-4472 to the chassis with the two screws attached for that purpose.
8.
Visually verify the installation. Make sure the device is not touching
other devices or components and is fully inserted into the slot.
9.
Plug in and power on the computer.
The NI PXI-4472 is now installed.
Testing the Device
The NI 447
X
is completely software configurable. The system software
automatically allocates all device resources, including base memory
address and interrupt level. This device does not require DMA controller
resources from the computer.
To check the configuration of the NI 447
X
, and to test its resource
allocations to be sure they do not conflict with any others, refer to the
DAQ Quick Start Guide
included with the device.
You can modify data acquisition-related settings, such as input coupling,
through NI application-level software, such as LabVIEW or Measurement
Studio, or with NI-DAQ. Refer to device configuration instructions in the