Figure 5. Puncturing and Inserting the Desiccant Pack
Puncture Here
Complete the following steps.
1.
Confirm that the power to the device has been turned off.
2.
Loosen the four captive screws on the door of the device.
3.
Open the door of the device slowly.
4.
Activate the desiccant pack by piercing the Mylar outer bag with the provided 1/16 in.
punch. See the figure above for where to puncture the desiccant pack.
5.
Place the desiccant in the clear pocket so that the pierced hole is not blocked by the
pocket, door, or Mylar folding on itself.
6.
Clean the gasket and gasket mating surface prior to closing the device door.
7.
Close the door slowly.
8.
Tighten the four captive screws on the door of the device to 1.3 N · m (11.5 lb · in.).
Network Connection
The first time you power up the MON-10496, it attempts to initiate a DHCP network
connection. If the MON-10496 cannot initiate a DHCP connection, it connects to the network
with a link-local IP address with the form
169.254.
x
.
x
. Refer to the NI InsightCM Help in
software or online at
for more information on finding and adding a
device to NI InsightCM.