Advanced System Calibration Settings
Refer to the following section for more information about configuring advanced system
calibration settings in the
NI RF System Calibration Assistant Configurator
window.
Note
NI recommends that you familiarize yourself with the following settings
before making changes to the default settings.
Generator Paths
The power meter measures one signal generator (SG) path. The
Generator Paths
window
provides an example of how the SG system calibration works. The SG path is connected to the
reference antenna to calibrate the path loss on the system.
You can add as many paths as your application needs for your specific configuration.
However, this manual contains information about configuring the SG paths for OTA only.
Generator Calibration Settings
The frequency spacing settings, including levels at which to calibrate, are selected by default.
For OTA, it is important to use same level as the level used in the
Analyzer Calibration
Settings
window to maintain the same RF state of the instrument.
Analyzer Paths
The analyzer measures two signal analyzer (SA) paths. The reference antenna (RA) values are
de-embedded so the SA path loss includes the loss between the end of the RA and the SA
instrument.
Analyzer Calibration Settings
The frequency spacing settings, including levels at which to calibrate, are selected by default.
For OTA, it is important to set the reference level low enough to get low variation.
Getting Started with Step-by-Step Position Sweeping
With step-by-step position sweeping, sweeping is exclusively controlled by the test execution
engine which consumes a list of predefined positions and sequentially performs a
measurement at each point.
Step-by-step position sweeping consists of the following typical process flow:
1.
Move the positioner to the desired position
2.
Set up your DUT
3.
Set up the instrument
4.
Take the measurement
5.
Repeat the setup and measurement for each position
Complete the following actions before getting started with step-by-step position sweeping.
•
Ensure that the system is properly configured as specified in
•
Remove the laser alignment fixture.
30
|
ni.com
|
mmWave OTA Validation Test Reference Architecture