Required Software for Instrumentation
The following instruments require specific driver software that is not installed with the
mmWave OTA Validation Test Software. Refer to the instrument documentation available at
ni.com/manuals
for more information about installing required driver software.
•
PXIe-5831 Vector Signal Transceiver and mmRH-5582 mmWave Radio Head
•
PXIe-6361 Multifunction I/O Module
•
Operating the Chamber
You must complete the following actions when operating the chamber.
Notice
The leakage current from the EMI AC filter inside the chamber may cause
certain residual current circuit breakers (RCCB) to trip. Use the external isolator
provided with the chamber if you are connecting the chamber to an RCCB.
Power On and Reset the Chamber
Complete either of the following actions after you power on the chamber.
•
If the key switch is set to Manual, press the Reset button.
•
If the key switch is set to Auto, close the chamber door, then press the Reset button.
Opening the Chamber Door During Operation
It is important that the chamber door remain closed when you are performing a test. If the key
switch is set to Auto and you open the door during operation, all movement in the chamber
stops and you must press Reset to return the chamber to the default state. No action is required
if the key switch is set to Manual and you open and close the door during operation.
Related Information
on page 5
Changing the Chamber IP Address
Notice
Changing the IP address is a low-level function of the internal controller in
the chamber and must only be changed if necessary. If the IP address is not updated
correctly, the chamber may not function properly.
You must use Internet Explorer to complete the following steps.
1.
On the client PC connected to the chamber, open Internet Explorer and enter
192.168.2.10
in the address bar.
The NI Web-Based Configuration and Monitoring utility opens.
2.
Click the
Login
button and enter
admin
in the
User name
field and leave the password
field blank.
3.
Change the system password:
a)
Click the
Security Configuration
button in the toolbar.
b)
Double-click
admin
in the list of users under the
Users
tab.
14
|
ni.com
|
mmWave OTA Validation Test Reference Architecture