10
|
ni.com
|
NI ISM-7400 User Manual
Connecting the Digital Output
The NI ISM-7400 has a digital output labeled OUT. This output closes to signal a fault condition.
Figure 13.
Internal Circuit Diagram
Use this output to drive LEDs, relays, and the inputs of other electronic devices like PLCs. The
positive collector and negative emitter terminals of the output transistor are available at the
connector. This allows you to configure the output for current sourcing or sinking.
Diagrams of each type of connection follow.
Caution
Do not connect the output to more than 30 VDC. The current through the
output terminal must not exceed 80 mA
Figure 14.
Sinking Output
Figure 15.
Sourcing Output