Chapter 3
Local Command Set
©
National Instruments Corporation
3-55
*CLS
488.2 Intent
Clear the device status data structures, and force them to the Operation Complete Query
Idle state.
Command Syntax
*CLS
Action
None.
*ESE
488.2 Intent
Set the GPIB-VXI/C’s Standard Event Status Enable (ESE) register bits.
Command Syntax
*ESE <byte value>
where
<byte value>
is the new value of the ESE register.
Action
Sets ESE to
<byte value>
.
Example
Set the ESE register to 45.
*ESE 45