Chapter 3
Local Command Set
©
National Instruments Corporation
3-85
TrigTickConf
Purpose
Configure the TIC chip internal dual 5-bit tick timers.
Query Syntax
TrigTickConf <mode>, <source>, <tcount1>, <tcount2>
where
<mode>
specifies the configuration mode.
and the value of
<source>
is the trigger line to configure as input to counter.
and the values of
<tcount1>
and
<tcount2>
are the number of input pulses (as a power
of two) to count before asserting TICK1 output or TICK2 output, respectively
(and terminating the tick timer if configured for non-rollover mode).
Value
Mode
0
Initialize the tick timers (rollover mode)
1
Initialize the tick timers (non-rollover mode)
2
Reload the tick timers leaving enabled
3
Disable/abort any count in progress
Value
Trigger Line
40 to 49
External source/destination (GPIO 0 to 9)
70
CLK10
71