Chapter 3
Local Command Set
©
National Instruments Corporation
3-59
*SRE?
488.2 Intent
Get the contents of the SRE register.
Query Syntax
*SRE?
Response
<SRE val>
<CRLF>
<SRE val>
is the current value of the SRE register. The default value is
FFh
.
*STB?
488.2 Intent
Get the contents of a device’s Status Byte.
Query Syntax
*STB?
Response
<STB value>
<CRLF>
where
<STB value>
is the current status of the path to the GPIB-VXI/C local
command parser.
*TRG
488.2 Intent
Cause a device to execute a stored trigger sequence.
Command Syntax
*TRG
Action
None.