Figure 33. Ambient Temperature Locations, Vertical Mounting Configuration
76.2 mm
(3.00 in.)
25.4 mm
(1.00 in.)
76.2 mm
(3.00 in.)
25.4 mm
(1.00 in.)
99.3 mm
(3.90 in.)
99.3 mm
(3.90 in.)
76.2 mm
(3.00 in.)
76.2 mm
(3.00 in.)
STATUS
IN
10 A TOTAL
9-30 V
FieldDAQ
OUT
0
10/100/1000
LINK/ACT
SYNC
1
Measure the ambient
temperature here
Measure the ambient
temperature here
Measure the ambient
temperature here
Measure the ambient
temperature here
Dimensions
For detailed dimensional drawings and 3D models, visit
and search for
FD-11634.
Mounting the Device Directly on a Flat Surface
For environments with high shock and vibration, NI recommends mounting the FieldDAQ
device directly on a flat, rigid surface using the mounting holes on the device. This mounting
technique requires four M5 or 10-32 panhead or sockethead cap screws appropriate for the
surface.
Complete the following steps to mount the FieldDAQ device directly on a flat surface.
1.
Prepare the surface for mounting the device using the surface mounting dimensions.
FD-11634 User Guide
|
© National Instruments
|
45
Содержание FD-11634
Страница 54: ...FD 11634 User Guide National Instruments 53 ...