Appendix B
Differences among the Lab-PC+, the DAQPad-1200, and the DAQCard-1200
DAQCard-1200 User Manual
B-2
ni.com
The following table summarizes the differences among the three products.
Functionality
Lab-PC+
DAQPad-1200
DAQCard-1200
Bus resource allocation
Jumpers
Software
Software (automatic)
Calibration
Potentiometers
Software
Software
DMA
Uses DMA
Cannot use DMA
Cannot use DMA
Waveform generation
Can do waveform
generation on one
DAC and
immediate updates
on the other DAC
Can do waveform
generation on one
DAC and
immediate
updates on the
other
Should do either
immediate update on both
or waveform generation
on both but not a
combination. Refer to
Chapter 4,
, for more
information.
FIFO size
512 samples
1,024 samples
1,024 samples
Maximum single-channel
acquisition rate
83.3 kS/s
100 kS/s
100 kS/s
Power-management modes
—
—
150 mA operating,
50 mA power-down mode
Fuse
Regular 250 mA
Regular 250 mA
Thermal 1.0 A