![National Instruments DAQArb 5411 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/national-instruments/daqarb-5411/daqarb-5411_user-manual_3428874019.webp)
Chapter 3
Signal Connections
DAQArb 5411 User Manual
3-2
©
National Instruments Corporation
ARB Connector
The ARB connector provides the waveform output. The maximum
output levels on this connector depend on the type of load termination.
If the output of a DAQArb 5411 terminates into a 50
Ω
load, the output
levels are ±5 V, as shown in Figure 3-2. If the output of DAQArb 5411
terminates into a high impedance load (HiZ), the output levels are
±10 V. If the output terminates into any other load, the levels are:
where V
out
is the maximum output voltage level, R
L
is the load
impedance in ohms, and R
O
is the output impedance on the
DAQArb 5411. By default, R
O
= 50
Ω
, but the software can also set it
to 75
Ω
.
Note:
Software will set the voltage output levels based on a 50
Ω
load termination.
For more information on waveform generation and analog output
operation, refer to Chapter 4, Arb Operation. For specifications on the
waveform output signal, see Appendix A, Specifications.
Figure 3-2.
Output Levels and Load Termination Using a 50
Ω
Output Impedance
V
out
=
±
x
10 V
R
L
R
L
+ R
O
ARB
DAQArb 5411
Load
±
5 V
R
L
=
50
Ω
ARB
DAQArb 5411
Load
±
10 V
R
L
=
HiZ
50
Ω
Load
High Impedance Load
R
O
=
50
R
O
=
50
Ω
Ω